Ultima 2Pplus - System specifications
Ultima 2Pplus
Laser input
- Coherent Chameleon Discovery NX TPC dual-output femtosecond laser
Widely tunable from 660 to 13200 nm
Fixed secondary output at ~1040 nm (>2.8 W)
Peak power >2.7 W
Laser power meter with pick-off window for monitoring
Integrated AOM modules for imaging and uncaging pathways
Switchable IR mirror mechanics for routing between galvo and resonant scanning paths
Laser scanner
- Dual independent sets of X-Y galvanometer mirrors for standard
Standard set for raster scan imaging
Second set for precise, programmable point scanning and photostimulation
Resonant scanner (8 kHz resonant galvo for up to 30 fps)
- Field of view:
Galvo scanning: 22.7 mm field number (1.11 × 1.11 mm with 16× 0.8 NA objective)
Overscan mode: 28 mm field number (1.4 × 1.4 mm with 16× 0.8 NA objective)
Scan modes: frame scanning, ROI scanning, freehand line scanning, scan rotation, optical zoom
Detectors
Two high-efficiency Hamamatsu GaAsP PMTs with mounting flanges
Pre-amplifier and HV controller optimized for GaAsP tubes
Close-proximity detector design for maximum photon collection with high NA objectives
Dual emission filters: ET525/70m-2p and ET595/50m-2p with t565lpxr dichroic beam splitter
2″ collection optics for maximizing photon throughput
Optics and nosepiece
Single objective nosepiece (M32 threading) with rotational axis for angled imaging (±87°)
Motorized Z-focus with reliable 250 nm step size (electronics capable of 100 nm increments)
Motorized pitch nosepiece for precise rotational control
Beam expander lens set (3×) for adapting to different objective back apertures
- Objectives supported
Nikon CFI Plan Apo Lambda 4×
Nikon CFI75 16×/0.8 NA
Nikon Apo 60× Water/1.0 NA
Epi-fluorescence imaging
Modified epi-illuminator arm with integrated LED excitation source
Widefield fluorescence imaging for sample positioning and overview imaging
Integrated CMOS camera for capturing epifluorescence images
Imaging chamber and enclosure
Light-tight Faraday cage enclosure (43.25″ × 36″ × 40″) with interlocked laser safety ports
Table laser safety covers (for 4′ × 8′ table)
Electronics
FPGA-based control electronics for precise synchronization
Multi-channel analog integration with 12-bit A/D conversion
Pre-amplifier and HV electronics tailored for GaAsP detectors
Remote controller for motorized Z-axis
Software
Prairie View scan control and acquisition software
Simultaneous acquisition from up to four channels with independent LUTs
Time Series acquisition for flexible experimental design
Resonant Galvo+ software module for high-speed imaging
Integrated calibration and drift compensation for ETL remote focusing modules
Available accessories
Operational protocols and risk assessments
The SOPs and RAs related to the use of the Ultima 2Pplus system are available https://drive.google.com/drive/u/0/folders/1biXgabX9BQkMIouT0Lx4Yg9P1y6UJ8bb, upon request. These documents are stored in the designated folder and include:
SOP for Ultima 2Pplus utilization
RA for Ultima 2Pplus with anesthesia