.. _Ultima-specifications: Ultima 2Pplus - System specifications ===================================== .. image:: ../_static/Ultima2Pplus.jpg :alt: *Ultima 2Pplus* :width: 1000px :align: center *Ultima 2Pplus* .. raw:: html Laser input ----------- - Coherent Chameleon Discovery NX TPC dual-output femtosecond laser - Widely tunable from 660 to 13200 nm - Fixed secondary output at ~1040 nm (>2.8 W) - Peak power >2.7 W - Laser power meter with pick-off window for monitoring - Integrated AOM modules for imaging and uncaging pathways - Switchable IR mirror mechanics for routing between galvo and resonant scanning paths Laser scanner ------------- - Dual independent sets of X-Y galvanometer mirrors for standard - Standard set for raster scan imaging - Second set for precise, programmable point scanning and photostimulation - Resonant scanner (8 kHz resonant galvo for up to 30 fps) - Field of view: - Galvo scanning: 22.7 mm field number (1.11 × 1.11 mm with 16× 0.8 NA objective) - Overscan mode: 28 mm field number (1.4 × 1.4 mm with 16× 0.8 NA objective) - Scan modes: frame scanning, ROI scanning, freehand line scanning, scan rotation, optical zoom Detectors --------- - Two high-efficiency Hamamatsu GaAsP PMTs with mounting flanges - Pre-amplifier and HV controller optimized for GaAsP tubes - Close-proximity detector design for maximum photon collection with high NA objectives - Dual emission filters: ET525/70m-2p and ET595/50m-2p with t565lpxr dichroic beam splitter - 2″ collection optics for maximizing photon throughput Optics and nosepiece -------------------- - Single objective nosepiece (M32 threading) with rotational axis for angled imaging (±87°) - Motorized Z-focus with reliable 250 nm step size (electronics capable of 100 nm increments) - Motorized pitch nosepiece for precise rotational control - Beam expander lens set (3×) for adapting to different objective back apertures - Objectives supported - Nikon CFI Plan Apo Lambda 4× - Nikon CFI75 16×/0.8 NA - Nikon Apo 60× Water/1.0 NA Epi-fluorescence imaging ------------------------ - Modified epi-illuminator arm with integrated LED excitation source - Widefield fluorescence imaging for sample positioning and overview imaging - Integrated CMOS camera for capturing epifluorescence images Imaging chamber and enclosure ----------------------------- - Light-tight Faraday cage enclosure (43.25″ × 36″ × 40″) with interlocked laser safety ports - Table laser safety covers (for 4′ × 8′ table) Electronics ----------- - FPGA-based control electronics for precise synchronization - Multi-channel analog integration with 12-bit A/D conversion - Pre-amplifier and HV electronics tailored for GaAsP detectors - Remote controller for motorized Z-axis Software -------- - Prairie View scan control and acquisition software - Simultaneous acquisition from up to four channels with independent LUTs - Time Series acquisition for flexible experimental design - Resonant Galvo+ software module for high-speed imaging - Integrated calibration and drift compensation for ETL remote focusing modules Available accessories --------------------- Operational protocols and risk assessments ------------------------------------------ The SOPs and RAs related to the use of the Ultima 2Pplus system are available https://drive.google.com/drive/u/0/folders/1biXgabX9BQkMIouT0Lx4Yg9P1y6UJ8bb, upon request. These documents are stored in the designated folder and include: - SOP for Ultima 2Pplus utilization - RA for Ultima 2Pplus with anesthesia