.. _SkyScan-specifications: SkyScan 1276 - System specifications ==================================== .. image:: ../_static/SkyScan1276.jpg :alt: *SkyScan 1276* :width: 1000px :align: center *SkyScan 1276* .. raw:: html X-ray source ------------ - Type: Tungsten source - Adjustable output: power Up to 20W and voltage between 40–100 kV - Spot size: < 5 µm at 4 W - Automated 6-position filter changer: no filter, Al 0.25mm, Al 0.5mm, Al 1mm, Al 1mm +Cu 0.03mm, low dose (dose reducing filter) X-ray detector -------------- - Sensor type: 16 megapixels cooled CMOS camera with coupling to scintillator by 1.8:1 fiber optic taper - Sensor dimensions: 4096 × 4096 pixels - Pixel size: 2.8 µm to 75 µm - Spatial resolution: 6 µm - Sensor binning: up to 8x (512 x 512 pixels) - Capability: Wide scan mode for large object - Readout rate: 4 frames/s full format and17 frames/s at 8x8 binning Imaging chamber --------------- - Temperature-controlled platform for maintaining physiological conditions - Integrated gas anesthesia system - Radiation Shielding: fully enclosed X-ray-safe chamber - Live Animal Monitoring: 5 MP color visible-light camera for real-time monitoring - Respiratory monitoring with breathing detection - Object capacity: up to 310 mm height and 75 mm diameter Rotation mechanics ------------------ - Adjustable rotation steps: from 0.1 to 1° steps - Rotation range: 180 and 360° option Software suite -------------- - SkyScan 1276 - data acquisition platform - NRecon - reconstruction and pre-processing - Dataviewer - Slice-by-slice inspection and registration tools - CTVox - High-quality volume rendering - CTAn - Advanced 2D/3D image analysis & quantification - CTVol - Surface model visualization and export (CAD/3D printing) Available accessories --------------------- - Hydroxyapatite phantoms for BMD calibration with a 2-32 mm diameter range - QRM phantom for spatial resolution calibration - Measurements calibration in Hounsfield Units (HU) - ECG patches for cardiac gating and physiological monitoring during in vivo imaging - Real-time dose meter inside chamber for dose tracking - Imaging cassettes of various diameters to accommodate ex vivo samples and whole-body scans of mouse and rats with optimized specimen positioning Operational protocols and risk assessments ------------------------------------------ The SOPs and RAs related to the use of the SkyScan 1276 system are available https://drive.google.com/drive/u/1/folders/1ZRjqZFXyPtIG-4KglzBfxUvnnrdUHj9P, upon request. These documents are stored in the designated folder and include: - SkyScan 1276 user manual - Radiation safety program manual - SkyScan 1276 workflow - SOP for SkyScan 1276 utilization - RA for SkyScan 1276 with anesthesia - Bruker useful information