.. _IVIS-specifications: IVIS Spectrum - System specifications ===================================== .. image:: ../_static/IVIS.jpg :alt: *IVIS Spectrum* :width: 1000px :align: center *IVIS Spectrum* .. raw:: html CCD camera ---------- - Back-thinned, back-illuminated Grade 1 CCD sensor - Thermoelectric cooling to -90°C for low dark current and noise - Pixel size: 13.5 µm - Minimum pixel resolution: 20 µm - Sensor dimensions: 2048 × 2048 pixels - Maximum sensor binning: 16 - Quantum efficiency: 85% in the 500–700 nm range and 30% in the 400–900 nm range Lens ---- - 6-inch diameter optics - Numerical aperture adjustable from f/1 to f/8 - Discrete magnifications: 1.5×, 2.5×, 5×, and 8.7× - Corresponding fields of view: 3.9 cm to 22.5 cm - High-resolution imaging down to 20 µm with small field of view Optical path ------------ - High-throughput imaging with a large field of view (up to 5 mice or 2 rats simultaneously) - 10 high-efficiency narrow bandpass excitation filters (415–760 nm, 30 nm bandwidth) - 18 high-efficiency narrow bandpass emission filters (490–850 nm, 20 nm bandwidth) - Optical switch enabling both epi-illumination and trans-illumination fluorescence modes Imaging chamber --------------- - Temperature-controlled platform for maintaining physiological temperature during imaging - Integrated RAS-4 gas anesthesia system - light tight chamber Software -------- - Radiometric calibration of measurements in radiance (bioluminescence) and radiance efficiency (fluorescence) for reproducible results - Spectral unmixing for separating multiple fluorescent or bioluminescent signals and background subtraction - Co-registration with anatomical atlases or CT/MRI for overlaying structural and functional data - 3D bioluminescence and fluorescence tomography with depth estimation and quantitative analysis - Exportation of DICOM image format for 3D co-registration Available accessories --------------------- - Fluorescence and bioluminescence phantom mouse for calibration, validation, quality control - Mouse Imaging Shuttle (MIS) for 3D multimodal co-registration with microCT or MRI systems Operational protocols and risk assessments ------------------------------------------ The SOPs and RAs related to the use of the IVIS Spectrum system are available https://drive.google.com/drive/u/1/folders/1yU5AkpK4jt7LfB_xHxhKC4837ZLwxuR2, upon request. These documents are stored in the designated folder and include: - IVIS Spectrum user manual - SOP for IVIS Spectrum utilization - SOP for RAS-4 rodent anesthesia system - RA for IVIS Spectrum with anesthesia - Revvity useful information